Slip rings for cT scanners & medical imaging

A CT gantry spins the X-ray tube and detector array at high speed on a ring roughly a metre across, while kilowatts of power go onto the rotor and multi-gigabit image data comes off it. Aggregate detector data is well above what sliding contacts can carry reliably, so data uses non-contacting optical or RF links while power stays on brushes. The system must also stay quiet, low-debris and electrically safe for patients.

Medical and laboratory

What the slip ring has to handle

  • Very large bore (patient aperture) with the slip ring wrapped around it
  • Several kilowatts of power transferred through contacting rings to the X-ray generator and rotating electronics
  • Non-contacting high-speed data link (optical or RF/capacitive) for detector data, typically 1-20+ Gbps aggregate
  • Bidirectional control channel between stationary and rotating gantry electronics
  • High surface speed at the ring circumference: gantry speeds of 120-280 rpm on a large diameter
  • Low wear debris, low audible noise and long maintenance-free life
  • Tight control of EMI from the data link and power rings, plus patient-safety insulation and leakage requirements

Typical figures

SpeedGantry rotation 120-280 rpm; slip rings run at 200+ rpm on a 1 m ring
BoreInside diameter up to 1270 mm (50 in)
DataNon-contacting data channels 1-20+ Gbps; single optical or RF channels typically 1.25-5 Gbps, spatially multiplexed optical links above 10 Gbps
PowerSeveral kilowatts through contacting power rings

Example figures from the published sources listed below. Your requirement may differ; an engineer will confirm against your duty cycle.

Recommended slip ring types

Standards to consider

  • IEC 60601-1 - general safety and essential performance of medical electrical equipment (insulation, leakage current, protection)
  • IEC 60601-1-2 - EMC requirements for medical electrical equipment, relevant to the gantry data link and power rings
  • IEC 60601-2-44 - particular requirements for X-ray equipment for computed tomography

The complete system

A slip ring rarely works alone. For cT scanners & medical imaging it is usually specified together with:

  • large-diameter gantry bearing
  • direct-drive or belt-drive gantry motor
  • servo drive
  • high-resolution encoder
  • contactless data link
  • brake

Specifying these together avoids mismatched connectors, cable routes and feedback interfaces, and gives you one point of contact for the whole axis.

Common problems to design out

  • Data link errors or EMI affecting image reconstruction
  • Brush wear debris contaminating rings and causing arcing or noise
  • Audible noise and vibration at high gantry speed
  • Contact resistance rise on power rings leading to heating

Series used in this application

Frequently asked questions

Why do CT scanners use contactless data transmission?

Detector data from a modern scanner typically exceeds 1 Gbps in aggregate, which sliding contacts cannot carry reliably. Optical or RF near-field links move the image data across the rotating gap while brushes are kept for power.

How fast does a CT gantry slip ring rotate?

Gantries commonly turn the tube and detector at roughly 120 to 280 rpm. Because the ring is around a metre across, the contact surface speed is very high, which drives brush material choice.

Optical or RF data ring - which is better for CT?

Optical links avoid high-frequency electrical signals on the large ring and can be fully isolated from the structure, which helps EMC. RF near-field links are usually slightly cheaper in electronics. Both typically run 1.25-5 Gbps per channel and can be multiplexed for more.

Sources

Spec builder

Configure a slip ring for cT scanners & medical imaging

Select every part you need, answer the questions an applications engineer would ask, and see your summary before we ask for any details.

  1. 1Products
  2. 2Performance
  3. 3Environment
  4. 4Volumes
  5. 5Your spec
Step 1 of 5

Engineer support

Let our experienced engineers do the hard work for you

Tell us what you need, whether it is this part, an equivalent or a complete axis built around it. An engineer will review your requirements and reply, normally within one working day.

  • Reviewed by an engineer, not an autoresponder
  • Like-for-like and better-fit alternatives
  • Standard, modified and custom options

Prefer to configure it? Use the spec builder.

We use your details to answer this enquiry and share them with the specialist engineering and supply partner best placed to fulfil it. See our privacy notice.

Ask an engineer